01.31
01.31
are designed for processing medium temperature measurement in methane conversion units under elevated oxygen concentration.
These units can replace thermocouples produced by OKAZAKI Manufacturing Company (Japan).
Sensors can be of explosion protection type 0ExiaIICT6 X or 1ExdIICT6 according to GOST 30852.10-2002. For further details see "Alternative designs & Ordering information" below.
Terminal heads can be equipped with transmitters with 4-20 mА DC output signal and/or digital signal via HART, PROFIBUS-PA, FOUNDATION Fieldbus protocol. It is recommended to use heads of 16, 23, 28 modifications for easy wire connection to transmitters. Terminal head can be equipped with standard or special-purpose cable gland. For further details see "Alternative designs & Ordering information" below.
Hot junction |
one or two |
Ungrounded |
|||
Vibration resistance GOST R 52931 |
group N2 |
||||
Nominal (rated) pressure |
4 MPa |
||||
Seismic resistance MSK-64 |
grade 9 for unit installation up to 70 m above zero mark |
||||
Climatic category GOST 15150 |
UHL2Ambient temperature: |
-60..+120ºС for general purpose products |
|||
-60..+85 ºС for Ex version with analogous signal |
|||||
-55..+85ºС for products with 4-20mА signal |
|||||
Calibration |
- GOST 8.338-2001 – for sensor fitting length larger than 250 mm, without measuring transmitters; |
||||
Response time τ 0,63 |
90 seconds |
01.32 |
|||
120 seconds |
01.31 |
Group of operating conditions |
Probability of |
Specified lifetime |
Mean |
Guarantee |
III |
0.95 for 8 000 hours |
1 year |
2 years |
1 year |
IV |
Not rated |
Not rated |
Not rated |
Not rated |
Measurement accuracy of sensors with 4-20 mА DC signal and/or digital signal via HART protocol are given taking into account accuracies of thermocouple itself, transmitter and transmitter's cold junction compensation.
Output signal |
Symbol |
Measurement accuracy |
Output |
Symbol |
Measurement accuracy |
4-20mА |
c0Т40 |
±0,4% ∙tn* or ±1,5°С |
4-20mА+HART |
c1N25 |
±0,25%∙ tn or ±0,9°С; |
c1Т50 |
±0,5%∙ tn or ±2,0°С |
c0N40 |
±0,4%∙ tn or ±1,2°С; |
||
c2Т80 |
±0,8%∙ tn or ±2,5°С; |
c1N50 |
±0,5%∙ tn or ±1,7°С; |
||
|
|
c2N80 |
±0,8%∙ tn or ±2,5°С; |
* - tn thermocouple range should be multiplied by specified value in в %. Select higher value.
CT type |
Thermowell material |
Group of operating conditions |
Application |
Calibration interval |
Mean lifetime |
CTK |
Т45 |
III |
-40…+1100 |
1 year |
2 years |
CTN |
Т45 |
||||
IV |
-200…+1250 |
Not rated |
Not rated |
CTK |
|
Exi |
01.31 |
— |
0 |
21 |
— |
c1 |
|
— |
U |
|
— |
Т45 |
— |
30 |
— |
L |
— |
997 |
1 |
|
2 |
3 |
|
4 |
5 |
|
6 |
7 |
|
8 |
9 |
|
10 |
|
11 |
|
12 |
|
14 |
NN |
Name |
Code |
Description |
|||||
1 |
Sensor type |
CTK, CTN |
Mineral insulated cable thermocouple with type according to IEC 60584 and ASTM E230 |
|||||
2 |
Explosion protection |
Not filled |
general purpose electrical equipment |
|||||
Exi |
0ExiaIICT6 X, intrinsically safe circuit |
|||||||
Exd |
1ExdIICT6, explosion-proof enclosure |
|||||||
3 |
Modification |
01.31, 01.32 |
|
|||||
4 |
Cable gland |
0 |
standard cable gland of terminal head |
|||||
А-Z |
Special cable gland (see table 3 page 1-13) |
|||||||
5 |
Sensor terminal |
15; 16 |
aluminium head |
IP66/IP68 |
1ExdIICT6 / 0ExiaIICT6 X |
|||
20; 22 |
aluminium head |
IP65 |
general purpose |
|||||
21; 23; 24; 25; 26; 28; |
aluminium head |
IP66 |
0ExiaIICT6 Х or general purpose |
|||||
6 |
Tolerance class |
c1; c2 |
For further details see table 5 page 2-9 |
|||||
7 |
Output signal, transmitter accuracy symbol, see table 5, page 2-10 |
Not filled |
analogous signal according to NSC |
|||||
Т40 |
4-20 mА |
for c0 |
||||||
Т50; Т70 |
for c1 |
|||||||
Т80; Т100 |
for c2 |
|||||||
H25 |
4-20 mА +HART |
Individual sensor calibration (c1) |
||||||
H40 |
for c0 |
|||||||
H50 |
for c1 |
|||||||
H80 |
for c2 |
|||||||
8 |
Sensor hot junction design |
G |
grounded junction, general purpose |
|||||
U |
ungrounded junction |
|||||||
9 |
Number of wire couples |
Not filled |
1 wire couple |
|||||
2 |
2 wire couples |
|||||||
10 |
Reinforcement material |
Т45 |
alloy CrNi45U |
|||||
11 |
Outer diameter |
17 |
mm for 01.32 |
|||||
30 |
mm for 01.31 |
|||||||
12 |
Probe length |
800÷1300 |
probe length L to the working end, mm |
|||||
13 |
Modification number |
994 |
flange |
"tongue" |
74 / 84 / 114 / 156 |
01.31 |
||
995 |
"male" |
84 / 114 / 156 |
||||||
996 |
"male" |
72 / 125 / 165 |
||||||
997 |
"male" |
87 / 135 / 175 |
||||||
998 |
"male" |
100 / 125 / 160 |
||||||
999 |
"male" |
75 / 125 / 160 |
||||||
993 |
fitting |
thread M20x1.5 |
01.32 |
CTK Exi 01.31-023 - c1 - U2 - Т45 - 30 – 1000 – 999 – temperature sensor based on cable thermocouple 01.31, modification version 999 with mounting flange 75/125/160 mm, tolerance class 1, two ungrounded hot junctions (U2), heat-resistant thermowell (Т45) 30 mm diameter, insertion length (L) 1000 mm. Explosion protection - "Intrinsically safe circuit ia", 0ExiallCT6 X.